Thera Lase Surgery equipment has the functions of emitting high power infrared LASER light (up to 4.5 W) used in surgery, and low power infrared LASER light (100mW) used in LASER therapy.
SPECIFICATIONS
* Supply voltage: 90-240V~
* Weight: 2,195 Kg
* Dimensions: 15 x 10 x 19,5 cm
Laser Therapy Function:
- Wavelength of the infrared laser: 808nm ± 10nm
- Useful power of the infrared laser: 100mW ± 20%
Surgery Function:
- Wavelength of the infrared laser: 808nm ± 10nm
- Useful power of the infrared laser: 4,5 W ± 20%
- Fiber diameter: 400µm or 600µm
- Exposure mode: single, continuous ou repetitive
- Single Exposure:
* Exposure time: 05 - 5000ms
- Repetitive Exposure:
Duration pulse: 05 - 95ms
Interval pulse: 05 - 95ms
Includes:
* Fiber optics pen for therapy (600 µm)
* Fiber optics pen surgery (fiber optics with 400µm or 600µm of diameters)
* Cap for fibre optics pen surgery
* Cap for fiber optics pen therapy
* Foot switch with cable
* Cover for laser output
* Interlock connector
* 01 spacer
* AC cable
* 02 fuses
* 02 protective goggles (user)
* 01 protective goggles (patient)
* 01 DVD
* Cleavage pen
* Stripper
* Transportation case
* User’s Manual
* Warranty